• Isotropic & True RMS measurement
  • Spectrum analysis probe
  • Measurements in accordance with International Standards

Specifications

  Electric Field Magnetic Field
Sensor type Isotropic patented electrodes Isotropic patented electrodes
Frequency range 1 Hz – 400 kHz 1 Hz – 400 kHz

Field Strength Mode

Measurement range 1 V/m to 100 kV/m 50 nT - 10 mT (100 Hz - 10 kHz)
Upper range increases linearly with decreasing frequency below 100 Hz.
Upper range decreases linearly with increasing frequency above 10 kHz.
Graphical display RMS, Axis Values, AVG, MAX, MIN, PEAK, RMS time graph
Peak value digital realtime digital realtime
Resolution < 0.4 mV/m above 8 Hz < 0.1 nT (at 50 Hz) and < 0.05 nT above 100 Hz
Noise level < 1 V/m (10 Hz - 400 kHz) < 50 nT (10 Hz – 400 kHz)
Typical Uncertainty (1) 0.67 dB 0.60 dB

Weigthed Peak Method mode

Measurement range 200 (min) 200 (min)
Graphical display PEAK (%), AXIS VALUES (%), AVG (%), MAX (%), MIN (%), RMS (%),Time graph
Standards/Limits EU Directive 2013/35/EU, FCC/IEEE, Safety Code 6, ICNIRP, BGV B11, Chinese Standard.
Easy software update to future modifications and to other limits.
Typical Uncertainty (1) 0.67 dB 0.60 dB

Applications

  • Industry
    Assessment of workers’ exposure to EM fields in all kind of manufacturing facilities.
  • Railway
    Measurement of EM fields in trains and in the railway environment with respect to human exposure.
  • Power grid
    Measurement of the exposure to EM fields at transformer stations and high voltage line.
Top