The GrayWolf PC-3016A can be used as a stand alone 6 channel Particle Counter by plugging it into a mobile computer or into a sophisticated WolfPack/Advanced Sense system for simultaneous connection with GrayWolf's DirectSense probes.
The PC-3016A combines particle counter and mass concentration readout in one instrument.
Size Range: | 0.3 - 10.0 μm+ (145 μm max into inlet) |
Channel Sizes: | 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm Optional: inquire about other size cuts (Particle Count only) |
Flow Rate: | 0.1 CFM (2.83 LPM) |
Counting Efficiency: | 50% @ 0.3 μm; 100% for Particles >0.45 μm (per ISO 21501-4) |
Laser Source: | Extreme Life Laser Diode with >20 Years MTTF |
Zero Count Level: | <1 Count / 5 minutes (per ISO 21501-4) |
Concentration Limits: | 4,000,000 Particles / ft3 @ 5% Coincidence Loss |
Calibration: | Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter |
Count Modes: | Automatic, Manual, Beep, Concentration, Cumulative/Differential, Mass Concentration |
Data Storage (on-board): | 3,000 Sample Records, Rotating Buffer (includes particle and environmental data, plus alphanumeric location and time) |
Communication Modes: | RS232 converted to USB (via included RJ-45 to USB cable) for PCs or Printers. RS232 to AdvancedSense, WolfPack or PPC w/GrayWolf cable/software. |
Touch Screen Display: | 3.5” (8.89cm, Color TFT) |
Sample Inlet: | Isokinetic Sampling Probe |
Sample Output: | Internally Filtered to HEPA Standards (>99.97% @ 0.3μm) |
Vacuum Source: | Internal Pump, Automatic Flow Control |
Size Range: | 0.3 - 10.0 μm+ (145 μm max into inlet) |
Channel Sizes: | 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm Optional: inquire about other size cuts (Particle Count only) |
Flow Rate: | 0.1 CFM (2.83 LPM) |
Counting Efficiency: | 50% @ 0.3 μm; 100% for Particles >0.45 μm (per ISO 21501-4) |
Laser Source: | Extreme Life Laser Diode with >20 Years MTTF |
Zero Count Level: | <1 Count / 5 minutes (per ISO 21501-4) |
Concentration Limits: | 4,000,000 Particles / ft3 @ 5% Coincidence Loss |
Calibration: | Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter |
Count Modes: | Automatic, Manual, Beep, Concentration, Cumulative/Differential, Mass Concentration |
Data Storage (on-board): | 3,000 Sample Records, Rotating Buffer (includes particle and environmental data, plus alphanumeric location and time) |
Communication Modes: | RS232 converted to USB (via included RJ-45 to USB cable) for PCs or Printers. RS232 to AdvancedSense, WolfPack or PPC w/GrayWolf cable/software. |
Touch Screen Display: | 3.5” (8.89cm, Color TFT) |
Alarms: | Internal Alarm Buzzer. Alarms on Counts, Low Battery, Sensor Failure |
Sample Inlet: | Isokinetic Sampling Probe |
Sample Output: | Internally Filtered to HEPA Standards (>99.97% @ 0.3μm) |
Vacuum Source: | Internal Pump, Automatic Flow Control |