• Isotropic & True RMS measurement
  • Spectrum analysis probe
  • Measurements in accordance with International Standards IEC/EN 62311

Specifications

  Electric Field Magnetic Field
Sensor type Isotropic electrode Isotropic 3 cm2 coils
Frequency range 1 Hz – 400 kHz 1 Hz – 400 kHz

Field Strength Mode

Measurement range 10 V/m to 400 kV/m 200 nT - 50 mT (100 Hz - 10 kHz)
Upper range increases linearly with decreasing frequency below 100 Hz.
Upper range decreases linearly with increasing frequency above 10 kHz.
Graphical display RMS, Axis Values, AVG, MAX, MIN, PEAK, RMS time graph  
Peak value digital realtime digital realtime
Resolution < 0.4 mV/m above 8 Hz < 0.3 nT (at 50 Hz) and < 0.15 nT above 100 Hz
Noise level < 10 V/m (10 Hz - 400 kHz) < 200 nT (10 Hz - 400 kHz)
Typical Uncertainty (10 Hz - 100 kHz) (1) 0.67 dB 0.60 dB

Weigthed Peak Method mode

Measurement range 200 (min) 200 (min)
Graphical display PEAK (%), AXIS VALUES (%), AVG (%), MAX (%), MIN (%), RMS (%),Time graph
Standards/Limits EU Directive 2013/35/EU, FCC/IEEE, Safety Code 6, ICNIRP 1998 workers, ICNIRP 2010 workers, BGV B11
Easy software update to future modifications and to other limits.
Typical Uncertainty (10 Hz - 100 kHz) (1) 0.67 dB 0.60 dB

Technical Specifications

  Electric Field Magnetic Field
FFT Mode
Measurement range 40 mV/m – 400 kV/m 2 nT - 50 mT (100 Hz - 10 kHz)
Upper range increases linearly with decreasing frequency below 100 Hz.
Upper range decreases linearly with increasing frequency above 10 kHz.
Graphical display Frequency analysis, total field and axis
SPAN (Resolution) 400 Hz (1 Hz) - 4 kHz (10 Hz) - 40 kHz (100 Hz) - 400 kHz (1 kHz)
Noise level < 40 mV/m < 1.8 nT
FFT 1024 point FFT  

General Specifications

Isotropy ± 5 % ± 4 %
Temperature deviation [typ. at 60 Hz]
(referred to 25 °C, 50 % relative humidity)
- 0.005 dB/°C (- 15 °C to 40 °C) - 0.003 dB/°C (- 15 °C to 25 °C)
+ 0.003 dB/°C (25 °C to 40 °C)
Damage level > 600 kV/m > 5000 mT up to 60 Hz
Damage level decreases linearly with increasing frequency above 60 Hz
Linearity ± 1 % (typ.)
± 2 % (max.)
Weight 125 g
Probe size 275 x 33 mm Ø

(1) Total, counting isotropy, temperature deviation, resolution, frequency response, linearity, repeatability.

Applications

  • IEC/EN 62311
    Household appliances and similar apparatus: Measurement methods for electromagnetic fields with regard to human exposure.
  • IEC/EN 62822
    Electric welding equipment: Assessment of restrictions related to human exposure to electromagnetic fields.
  • IEC/EN 62233
    Assessment of electronic and electrical equipment for which no dedicated product- or product family standard regarding human exposure to electromagnetic fields applies