The TM data II utilizes the Tyndall effect for scattered-light measurements. Maximum intensity is achieved for particles with dimensions comparable to the wavelength of the light source. As this technique is insensitive towards coarse particles, no pre-separation is necessary. With no pump required, the instrument does not affect the aero-dynamic and concentration situation at the measurement site.
Principle of operation: | 1 Scattered-light measurement (Tyndall effect) |
Measured particle range: |
Respirable dust acc. to Johannesburg Convention, BMRC/ACGIH curve, max. sensitivity for particles of 1 μm aerodynamic diameter |
Calibration method: | Comparis on with gravimetric measurement |
Measurement volume: | approx.0.5cm³ |
Measurement range [mg/m³,calibrated]: |
0-100 (DEHS particles,d=1μm) |
Resolution [mg/m³,calibrated]: |
0.01 (DEHS particles,d=1μm) |
Detectionlimit[mg/m³]: | 0.01 (measurement time atleast 1min. and with exact zero-point adjustment) |
Linearity: | 1% |
Temperature dependence: | approx. 5% between 10 °C and 40 °C |
Measurement modes: |
|
Memory capacity: | atleast 80.000measurementvalues |
Interfaces: |
|
User interface: | Touchscreen, resistive (8.1cm/3.2”) |
Charging time: | approx.6h |
Operating time when fully charged: |
approx.8h |
Dimensions[mm]: | 201 x 101 x 51(Lx W x H) |
Weight: | approx.600g |
USB power supply: | 100V ...240 V AC (1 Amax.), 50...60Hz, Output: 5V /2,1 A(10,5Wmax.), Adapterset EU/GB/US/CN |
Min. system requirements: | PC with Windows7or8.1,dual-core processor, 2 GB RAM, harddisk with 200MB free memory, 1 free US Bor Bluetooth interface |
Accessories (optional): |
|